==Cranial Nerves==
==Cranial Nerves==
For IONM involving the cranial nerves, the purpose of electrical stimulation is to identify the location of the nerves in the tissue. With that knowledge, the surgeon can then avoid making contact with them.
For IONM involving the cranial nerves, the main purpose of electrical stimulation is to identify the location of the nerves in the tissue. With that knowledge, the surgeon can then avoid the area and the possibility of injuring the nerve.
#Facial Nerve.         
#Cranial Nerve VII. The facial nerve has five main branches that innervate the muscles of the face and mouth. The Temporal Branch innervates the muscles above the eye, including the          
#Cranial Nerve III IV and VI
#Cranial Nerve III IV and VI
#Other Cranial Nerves
#Other Cranial Nerves

Revision as of 06:20, 28 September 2019

Electromyography is the recording of electrical activity from muscle fibers and represents an important diagnostic tool to monitor nerve root function and motor neuron activity. Motor neurons generate action potentials during voluntary movement, when electrically stimulated, and in response to disease or injury. An electromyograph can detect these changes in electric potential.

The Neuromuscular Junction

Lower motor neurons from the anterior horn of the spinal cord and from the motor nuclei in the brainstem send axonal projections to the periphery where they form synaptic connections with multiple muscle fibers. This interface between an axon terminal of the motor neuron and a single muscle fiber is known as the neuromuscular junction or motor end plate. The lower motor neurons release acetylcholine as a neurotransmitter. Acetylcholine binds to nicotinic acetylcholine receptors located on the cell membrane (or sarcolemma) of the muscle fiber, a process that can lead to depolarization and muscle contraction.

Electromyography Recording

EMG activity can be recorded using different types of electrodes, including monopolar needles, concentric needles, bipolar needles, and single-fiber needles. Without the use of high and low frequency filters, EMG signals would be very noisy and difficult to interpret. The low frequency filter should be set to 10-30 Hz and the high frequency filter to 10-20 kHz, for example.

  1. Spontaneous EMG. After the recording electrodes are inserted into the muscle tissue, the background EMG activity is stable and quiet under healthy conditions. However, the presence of an injury or a pathology can generate spontaneous EMG activity in the motor neuron or post-synaptically at the level of the muscle fiber. Examples of spontaneous activity arising from the muscle fiber include fibrillation potentials, positive sharp waves, myotonic discharges, and complex repetitive discharges. Examples of spontaneous activity arising from the motor neuron include neuromyotonic tonic discharges, myokymic discharges, and tremors.
  2. Stimulated EMG. Electrically stimulated EMG activity is used in IONM to monitor motor neuron or cranial nerve excitation by determining the stimulus threshold required to elicit a compound muscle action potential (CMAP). A CMAP represents the simultaneous activation of multiple muscle fibers and is often characterized by a large amplitude, polyphasic waveform. To determine the stimulus threshold for a CMAP, we slowly ramp up the electrical current until the CMAP is present. The stimulus threshold is the minimum current needed to observe this response above the background noise.
  3. Single-Fiber EMG. Developed in the 1960's by Stalberg and Eskedt (citation), single-fiber EMG involves the use of small-surface recording electrodes to record the electrical activity of individual muscle fibers. The technique has proven to be useful in the diagnosis of myasthenia gravis and other neuromuscular disorders (Baruca et al., 2016). Single-fiber EMG measures the interval between action potentials from different fibers of the same motor unit, also known as 'jitter'. The measurements are usually performed in awake patients, during which time the patient is asked to contract a muscle while the clinician records EMG activity.

Peripheral Nerves

Cranial Nerves

For IONM involving the cranial nerves, the main purpose of electrical stimulation is to identify the location of the nerves in the tissue. With that knowledge, the surgeon can then avoid the area and the possibility of injuring the nerve.

  1. Cranial Nerve VII. The facial nerve has five main branches that innervate the muscles of the face and mouth. The Temporal Branch innervates the muscles above the eye, including the
  2. Cranial Nerve III IV and VI
  3. Other Cranial Nerves

Intraoperative Monitoring

  1. Pedicle screws
  2. Peripheral Nerves
  3. Brachial Plexus
  4. Dorsal Rhizotomy
  5. Skull Base Tumor
  6. Others

Anesthesia and Other Factors

  1. Anesthesia Methods
  2. Muscle Relaxants
  3. Temperature
  4. Tourniquet
  5. Others

References